Abstract
Possible extended utilization of a 68Ge/68Ga source as a surrogate 18F reference source was studied. Some effects and uncertainties might arise from the difference of constructions between chambers that were used in the source calibration and those practically in use. To extend utilization of the surrogate source, response ratio of real 18F to 68Ge/68Ga sources were determined for three models of chambers. Monte-Carlo simulation using EGS5 code was also carried out to evaluate uncertainties expected. The present result shows that such a surrogate source would adapt to calibration even for chambers have with different construction from the reference chamber used as the determination of 18F equivalent activity of the 68Ge/68Ga surrogate source.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.