Abstract

This study investigates the critical significance of anode material selection in defining the energy spectrum and properties of X-ray photons in medical physics applications. Using the GATE platform and Monte Carlo simulations, a direct relationship between anode material atomic number and photon fluence is demonstrated. As the atomic number increases from Z = 29 (Copper) to Z = 74 (Tungsten), photon fluence rises by 62 %, indicating a substantial impact on X-ray production. Furthermore, the X-ray spectrum is affected by this material-driven changes, revealing a noticeable shift towards higher energy values: the mean energy of the continuous spectrum rises from 46.97 keV for Copper to 49.0 keV for Tungsten. The thermal properties of the material affect the temperature increase at the focal point. Rhodium and Molybdenum have a higher temperature rise than Copper (Cu) and Tungsten (W), because Cu and W have a greater thermal diffusion compared to other materials. These findings underscore the significance of anode material choice in optimizing X-ray systems which may enhance diagnostic accuracy and efficiency in diverse applications.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.