To evaluate the efficacy and safety of a new hydrodynamic catheter for removal of chronic iliac vein thrombus. Unilateral iliac vein thrombosis was induced in seven pigs by combining permanent coil and temporary balloon occlusion. Thrombectomy was performed with a new hydrodynamic catheter (10 F S.E.T.) 3 days after thrombus induction. After thrombectomy, the animals were killed and the iliac veins were examined histologically. Complete thrombectomy (100% thrombus removal) was achieved in three of seven animals, 75% removal in three of seven animals, and only 30% removal in one animal. The average thrombus removal was 75%. Successful re-establishment of flow was achieved in five of seven cases. Histologically, the thrombi were partially organized, meeting the histologic criteria for early chronic venous thrombosis. Minor venous wall damage caused by the thrombectomy procedure without acute hemodynamic consequences was observed in four of seven cases. The 10-F S.E.T. catheter was reasonably effective in removing chronic iliac vein thrombus with no hemodynamically significant complications.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.