Abstract
An experimental thermoradiotherapy study was started in 1986. For this study a hyperthermia system was developed for the heating of human tumours xenotransplanted into nude mice. Our treatment device was a four-channel computer-controlled hyperthermia system. Temperature was monitored by microwave radiometry at 3 GHz. Specifications of the radiometer were first evaluated under reference conditions, then thermal dosimetry was studied using non-invasive measurement of the brightness temperature TOR. Hyperthermia treatments were simulated in phantom material with radiometric monitor values TR of 40 and 41. In conjunction with the parameters of brightness temperature TOR, surface temperature TO and water bolus temperature TW thermal modelling was performed. Finally, we studied the influence of a perfused phantom to the microwave thermometry.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: International journal of hyperthermia : the official journal of European Society for Hyperthermic Oncology, North American Hyperthermia Group
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.