Abstract

The low limit of detection (LLD) plays an important role in biomolecular assays, especially for early-stage assays. Biomolecular detections usually involve the use of two main elements: a reagent and an analyzer, which both greatly contribute to the LLD. In this work, the relationships among the LLD and reagent-related factors are investigated. The to-be-detected biomolecule is c-reactive protein (CRP) as an example. The assay method is immunomagnetic reduction (IMR). The components of reagent are Fe(3)O(4) magnetic nanoparticles bio-functionalized with antibodies against CRP, dispersed in pH-7.4 phosphate buffered saline solution. Several key factors of the reagent, such as particle concentration, volume ratio of reagent to sample, and particle size, are manipulated to optimize the LLD of detecting CRP.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.