The purpose of this study was stimation of optimal percentage of lateral uncoverage of the acetabular component during total hip arthroplasty for patients with severe developmental hip dysplasia. Mathematical computer modeling based on the finite element technique and the mechanical experiment were performed. Critical values of uncoverage enabling safe primary fixation of acetabular component were estimated in designed models. Using the finite element technique and the mechanical experiment on pelvis models, a possibility of mounting an acetabular component with moderate uncoverage within 25% without screws and with significant uncoverage to 35% with an additional two-screw fixation was demonstrated. This study provides additional guidance on optimal acetabular uncoverage assessment and fixation methods of surgeons performing THA on patients with DDH.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.