Abstract

For the tritium assay at the Tritium Laboratory Karlsruhe (TLK) which started operation in Nov.94 an isothermal shield calorimeter was built. The calibration of the calorimeter was performed with an electrically heated simulator covering the range between 1 mW and about 2 W. After the calibration tritium activities in four Amersham uranium getter containers were determined. The excellent reproducibility and high precision of the measurements with the KfK-calorimeter was shown. The tritium activity inventories could be determined with relative standard deviations from the average value of the order of 0.01% to 0.1% depending on the activity level. These results demonstrate the suitability of the KfK-calorimeter as a nondestructive method for a careful assay of tritium. 3 refs., 11 figs., 2 tabs.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.