Abstract
The aim of this study has been the experimental determination of the energy dependent dose-to-water response of TLD600 and TLD700 thermoluminescent detectors (Harshaw) in X-ray beams with mean photon energies from about 20 to 200keV in comparison with 60Co gamma rays and 6MV X-rays. Experiments were carried out in collaboration with the German secondary standard laboratory PTW Freiburg. The energy dependent relative responses of TLD600 and TLD700 thermoluminescence detectors were determined at radiation qualities between 30kVp and 280kVp. The overall uncertainty of the measured values was characterized by standard deviations varying from 1.2 to 3%. The present results agree with previous studies on the energy dependent dose-to-water response of TLD100. As an application example, the results were used to measure doses associated with X-ray imaging in image-guided radiotherapy.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.