Abstract

Liquid electrode plasma atomic emission spectrometry (LEP-AES) is a novel elemental analysis method where a sample solution is placed in a small cuvette that has a narrow microchannel at the center. High voltage is applied to the solution from both ends of the cuvette channel, and then microplasma is generated in the channel. The microplasma is employed in atomic emission spectrometry as a plasma source. Excitation temperature was deduced using a Boltzmann plot. The temperature was determined to be 6200 K with a plastic chip at an applied voltage of 800 V. To investigate effects of the experimental conditions on excitation temperature, applied voltage, type of acid, and the shape of the cross section of a microchannel at the center of the cuvette were varied. The excitation temperature was not influenced by applied voltage or type of acid, but was affected by the shape of the cross section of the microchannel.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.