Abstract
Ex vivo testing techniques were used to determine the ferromagnetic qualities of, presence of heating in, and artifacts produced by 13 different heart valve prostheses exposed to a 1.5-T (64-MHz) magnetic resonance (MR) system. None of the heart valve prostheses showed a measurable deflection in the 1.5-T static magnetic field. Only minimal artifacts were produced during MR imaging with a fast spoiled GRASS (gradient-recalled acquisition in the steady state) pulse sequence. The largest temperature changes measured during a "worst case" MR imaging sequence (estimated average specific absorption rate, 2.5 W/kg; estimated spatial peak specific absorption rate, 7.6 W/kg) were +0.2 degree C with the implant imaged "in air" and +0.3 degrees C with the implant imaged in normal saline. Therefore, MR procedures performed with a 1.5-T (64-MHz) MR system may be performed safely in patients with any of the 13 different heart valve prostheses evaluated in this study.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.