Abstract

An industrial type 2MeV/20kW ILU-6 electron beam accelerator is operational in India for development of applications and technology demonstration to the Indian Industry in the field of polymer modifications (cross-linking and degradation). In order to adjust the treatment conditions and to control the good operation of the accelerator, it is necessary to study dose response as a function of various operational parameters. An experimental procedure for calibrating Cellulose Triacetate (CTA) film strip dosimeters in electron beam using total absorption graphite calorimeters is discussed and compared it with the results obtained from gamma calibration. Dosimetry data for process application, where the irradiation target is plane polymer sheet, have been obtained under various operational parameters.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.