Abstract

To evaluate ultrasound transmission velocity (UTV) for assessment of mechanical bone quality, an ex vivo comparison of different bone types measured with UTV, 2-dimensional (2D) histomorphometry and with 3-dimensional (3D) radiology (cone-beam computerized tomography [CBCT], computerized microtomography [μCT]) was conducted. Clinical cortical, cancellous, and mixed bone (each n = 6) was measured via UTV (m/s), CBCT (white pixel/black pixel ratio [WP/BP]), μCT (bone volume/total volume [μBV/TV]), and histomorphometry (bone volume/total volume [hBV/TV]). UTV values were correlated with 2D-histomorphometry and 3D-radiologic results. For the cortical, cancellous, and mixed bone samples, respectively, UTV values were 1,945.17, 1,266.9,and 1472.2 m/s, WP/BP quotients were 0.96, 0.15, and 0.33, μBV/TV quotients were 0.94, 0.2, and 0.47, and hBV/TV quotients were 0.96, 0.24 and 0.39. Significant correlations between UTV and the other methods were seen (P < .0001). Similar to the other methods, UTV is able to discriminate between different bone types ex vivo.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.