Abstract

This study evaluated the sensitivity of USP Prednisone tablets (Reference Standard Lot P0E203) to dissolved gas in the medium in USP Apparatus 2. This lot will be introduced into commerce over the summer of 2006 as a replacement for current Lot O0C056. The effect of deaeration techniques and water purification methods on dissolution was also studied. Dissolution tests were performed in water containing nine different concentrations of dissolved gas. The oxygen and total dissolved gas pressure was monitored using a dissolved gas and oxygen meter. Other experimental variables such as the equipment,shaft position,and vessel position were kept constant throughout the study. The same analyst performed all studies. The study tablets were highly sensitive to dissolved gas in USP Apparatus 2. An increase in both the amount of prednisone dissolved and the variability of results was observed when the dissolved oxygen concentration exceeded 6 mg/L. The corresponding sum of the partial pressure of gases other than oxygen and the percent saturation were 510 mm Hg and 68%,respectively.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.