Abstract

The aim of this study is to measure the dose area product (DAP) in digital radiography by using a DAP meter to determine the X-ray exposure. Pediatric X-ray examinations can be obtained for any radiographic examinations using the selected radiographic examination parameters (kVp and mAs), the DAP information recorded. The best peak signal-to-noise ratio (PSNR) at a fixed tube voltage of 70 kVp was obtained at tube currents of 20 and 32 mA, whereas the best PSNR at a fixed tube current of 25 mA was obtained at a tube voltage of 73 kVp. The fixed tube voltage of 70 kVp and the fixed tube current of 25 mA could help to obtain the best image quality and depict the spatial resolution of an anthropomorphic torso phantom radiographic examination. The normalized data over the DAP were provided to determine the patient dose from radiography.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.