Abstract

The Hematoflow (Beckman Coulter, USA) is a new automated hematology analyzer, which provides a 16-part white blood cell count (WBC) differential. We evaluated the differential WBC count performance of the Hematoflow. 101 blood samples from patients were selected for comparison analysis. The methodological comparison of the WBC differential parameters of neutrophils, lymphocytes, monocytes and eosinophils showed good correlations among 4 different analyses. More than 1% of blast cells were counted in 30 of 101 samples. A good correlation for blast cell counts obtained by Hematoflow was found with the reference manual method (r=0.9637, P 0.0001). For blast B, Hematoflow shows good correlation with reference method results but did not identify blast T. These results demonstrate that the Hematoflow has a comparable performance with the Sysmex XE-2100 and indicate that B cell lineage ALL can be identified by the use of the Hematoflow in an initial evaluation of acute leukemia.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.