Abstract

The Bureau International des Poids et Mesures (BIPM) has established a measurement chain allowing calibration of capacitance standards in terms of the quantized Hall resistance (QHR). An important element in the chain is a quadrature bridge linking a pair of ac resistors of values 2R/sub K/ /spl ap/ 51.6 k/spl Omega/ to a pair of capacitance standards. The quadrature bridge can be operated at five different frequencies: 513, 1027, 1541, 3082, and 6164 Hz. For such measurements, we use different ratios (1/1, 4/1 and 1/4) for the main inductive voltage divider in the quadrature bridge and three different pairs of capacitors of values 3000, 2000, and 1000 pF. A calculable coaxial resistance of 1290.6 /spl Omega/ (R/sub K//20) is used as a reference to evaluate the frequency dependence of the 51.6-k/spl Omega/ resistances. This allows the calibration of capacitance standards at the five different frequencies. The measured frequency dependences of 10 and 100 pF capacitance standards are reported.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.