Abstract

The flow cytometric crossmatch is a technique that is increasingly being used by clinical transplant laboratories. In this multicenter study by the British Society for Histocompatibility and Immunogenetics Flow Cytometry Group, a series of crossmatches were carried out to determine whether different centers obtained same results when performing the same crossmatch. There was greater than 80% agreement among participating laboratories on the results of 35/54 tests. There was no clear agreement in the remaining 20 cases. Quantitative analysis, estimating the number of cell-bound fluorescein molecules, demonstrated that differences in the criteria used by each center to define a positive crossmatch were responsible for some discordant results. When applied, definition of positivity based on the molecules of fluorescein increased concordance from 57.5% to 81.4%.l. These results suggest that a criterion for the interpretation of results based on quantitative analysis of bound antibody may be more reliable than methods in current routine use.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.