Abstract

To improve time to identification of pathogens and detection of resistance genes, we evaluated the BioFire FilmArray Blood Culture Identification Panel (BCID) as compared to: (1) direct MALDI-TOF MS (DM) and (2) standardized culture-based identification (ID) with antibiotic susceptibility testing (AST). BCID gave an accurate identification in 102/112 (91 %) of cases (102/103 for on-panel organisms). DM gave an accurate identification in 91/112 (81 %) of cases, with 13/91 (14 %) requiring repeat testing from the residual pellet. The mean time to an identification result was 2.4 and 2.9 h for BCID and DM, respectively. Standardized ID and AST results were available at a mean time of 26.5 and 33 h, respectively. There were 44 BCID/DM results that had an antimicrobial treatment change made based on rapid identification and resistant gene detection of pathogens. Both BCID and DM are accurate and rapid methods for the identification of new positive blood culture pathogens.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.