Abstract
Performance characteristics of the COBAS hepatitis C virus (HCV) TaqMan analyte-specific reagent (TM-ASR) assay using the QIAGEN BioRobot 9604 for RNA extraction were evaluated and compared to the COBAS Amplicor HCV Monitor V2.0 (Amplicor) and Versant HCV bDNA 3.0 (Versant) assays using clinical samples. Calibration of TM-ASR using Armored RNA allowed determination of the distribution of HCV RNA in clinical samples, using 22,399 clinical samples. Limit of detection, linearity, and inter- and intraassay assay precision were determined for the TM-ASR assay using multiple clinical specimen panels across multiple determinations. Genotype specificity for the TM-ASR assay was determined using samples with different HCV RNA genotypes evaluated and compared against predetermined results. Contamination control of the TM-ASR assay was evaluated using pools of HCV RNA-positive and -negative samples tested in a checkerboard pattern over 12 runs of 96 samples. Correlation of the TM-ASR, Amplicor, and Versant assays was determined using 100 paired clinical samples and Deming regression analysis. The TM-ASR performed well with respect to linearity, precision, and contamination control. The correlation between TM-ASR and the Amplicor and Versant assays was poor, with large differences between assay results for individual samples. Calibration of the TM-ASR assay with Armored RNA allowed for a wide dynamic range and description of the distribution of HCV RNA in clinical samples.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.