Abstract

An interlaboratory proficiency testing programme for melamine in milk was organized for field laboratories in Hong Kong, China, during the melamine crisis in late September 2008. One blank test sample and three homogenous samples prepared by gravimetric spiking of melamine at the concentration range of zero to 4.5 mg kg−1 were given to participants in this programme. A total of 13 participants returned the results to the organizer and they used either liquid chromatography-tandem mass spectrometry (LC-MS/MS) or gas chromatography-mass spectrometry (GC-MS) for their determinations. The performance of the participants was assessed by determining z-scores, calculated from the bias from the assigned reference values and Horwitz standard deviation. The median values of pooled data were found to be in good agreement with the reference values and the majority of the participants demonstrated their capabilities in the quantitative measurement of melamine in milk samples. However, four participants gave false-positive results for the blank test sample, probably due to cross-contamination from other samples, and they were requested to investigate the actual causes. In summary, eight participants (or 62%) demonstrated their competence for all the four test samples.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.