Abstract

Purpose: In order to measure spatial resolution of a PET tomograph in clinical conditions, this study describes and validates a method based on the recovery coefficient, a factor required to compensate underestimation in measured radioactivity concentration for small structures. Methods: In a PET image, the recovery factors of radioactive spheres were measured and their comparison with simulated recovery coefficients yielded the tomographic spatial resolution. Following this methodology, resolution was determined in different surrounding media and several conditions for reconstruction, including clinical conditions for brain PET studies. All spatial resolution values were compared with those obtained using classical methods with point and line sources. Results: In each considered condition, spatial resolution of the PET image estimated using the recovery coefficient showed good agreement with classical methods measurements, validating the procedure. Conclusion: Measurement of the recovery coefficient provides an assessment of tomographic spatial resolution, particularly in clinical studies conditions.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.