Abstract
ObjectiveTo evaluate the performance of newly developed glycated hemoglobin (HbA1c) clinical analytic reagents and HPLC columns, applied on Tosoh HLC-723 G8 Analyzer. MethodsNewly developed reagents and columns were used on a Tosoh HLC-723 G8 Analyzer (standard mode) system to measure both of qulity contorls and the clinical blood samples to evaluate the performances of these newly developed prodcuts including precision, accuracy, linearity, carryover, bias evaluation, correlation with commercial reagents, and stability according to CLSI recommendations. ResultsThe CV of intra-assay precision and inter-assay precision of quality control and clinical blood sample assays using Lirimax products were both less than 3.00%. And the REs of accuracy were less than 6.00%. Linearity: R2 = 0.9993 in the concentration range 4.77%–14.67%. Carryover: 0.05%. The Bland-Altman mean difference: −0.003583% HbA1c (CI: 0.07398: −0.08115); Passing-Bablok regression: y = 1.0022(0.9984:1.006)x-0.01097(-0.03776: 0.01582), R2 = 0.9996. Stability evaluation was also acceptable. ConclusionThe performance of newly developed products was well evaluated for HbA1c measurement on a TOSOH G8 Analyzer which shows excellent suitability for clinical assay.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.