Abstract

The current investigation was conducted to evaluate the accuracy of the Tanita 300WA leg-to-leg (LL) bioelectrical impedance analyzer (BIA) for measuring body composition in an American high-school-aged population. Body composition was determined in 77 students, comparing BIA measurements with the criterion, hydrostatic weighing (HW). Among the males, there were no significant differences found in percent body fat (%BF) between BIA (14.1 ± 7.8%) and HW (14.9 ± 9.1%); however, skinfolds (SK; 11.8 ± 7.2%) were significantly different from HW. A significant correlation in fat-free mass (FFM) was found between BIA and HW (r = 0.96, p < 0.001), and the standard error of estimate (SEE) for FFM was 3.28 kg for males. In females, a significant difference (p < 0.001) in %BF was found for both BIA (26.4 ± 5.7%) and SK (27.9 ± 5.1%) when compared with HW (23.6 ± 5.9%). The correlation in FFM between BIA and HW was lower (r = 0.78, p < 0.001) and the SEE for FFM was 2.93 kg for females. The Tanita 300WA LL-BIA system is appropriate for assessing body composition in male adolescents, but it warrants future research in female adolescents.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.