Abstract
The interpretation of measurements of bone strontium in vivo using energy dispersive x-ray fluorescence spectroscopy is presently hindered by overlying skin and soft-tissue absorption of the strontium x-rays. The use of imaging technologies to measure the overlying soft-tissue thickness at the index finger measuring site might allow correction of the strontium reading to estimate its concentration in bone. An examination of magnetic resonance (MR), computed tomography (CT) and high-frequency ultrasound (US) imaging technologies revealed that 55 MHz US had the smallest range of measurement uncertainty at 3.2% followed by 1 Tesla MR, 25 MHz US, 8 MHz US and CT at 4.3, 5.4, 6.6 and 7.1% uncertainty, respectively. Of these, only CT imaging appeared to underestimate total thickness (p < 0.05). Furthermore, an inter-study comparison on the accuracy of US measurements of the overlying tissue thickness at finger and ankle in nine subjects was investigated. The 8 MHz US system used in prior in vivo experiments was found to perform satisfactorily in a repeat study of ankle measurements, but indicated that finger thickness measurements may have been misread in previous studies by up to 17.7% (p < 0.025). Repeat ankle measurements were not significantly different from initial measurements at 2.2% difference.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.