Abstract

In this study, five different HPLC columns were heated to 200 °C using a homemade heating system which can be operated in temperature programmed mode. The column bleed as an indicator of induced degradation of the stationary phase material was evaluated using a charged aerosol detector (CAD) and an ultraviolet diode array detector (UV-DAD) at different wavelengths. The silica based C-18 stationary phase gave the highest bleed, and the carbon clad titanium dioxide column the lowest bleed. This was independent of both the detection technique and the wavelength.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.