Abstract

We evaluated the effect of airflow and gas composition on the linearity of measurement of airflow by a new disposable flowmeter. The flowmeter is based on the principle of differential pressure measurement across two symmetrically disposed Pitot tubes. Nonlinearities arising from the pressure-to-airflow relationship and sensitivity to changes in gas density were linearized with appropriate software and monitoring of the gas composition. With room air used as the respired gas, the measured tidal volume from a piston pump assembly was consistently within 1-2% of the target tidal volume for each of five flowmeters tested across physiological ranges of flow. Changing gas densities by varying concentrations of O2, CO2, and N2 led to errors in tidal volume measurement that ranged up to 6-8%. However, because the errors were predictable, they were corrected by software to within 0.6% of the target volume. Measurement of minute ventilation during exercise was within 1-2% of that determined from bag collections. We conclude that this type of flowmeter can accurately measure exercise minute ventilation and has advantages over some other flowmeters because of its ruggedness, reproducibility, and ease of sterilization or replacement compared with other flowmeters.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.