Abstract
Abdominal sound speed tomographic imaging using through-transmission travel time data on the body surface was investigated. To this end, a hundred kHz range low-frequency wave was used to reduce the wave attenuation within an inner body medium. A method was investigated for the reconstruction of the image with the smallest possible number of path data around the abdominal surface. Specifically, the data from a strong scattering spinal cord should be avoided. To fulfill the requirement, the smoothed path algebraic reconstruction technique was introduced. The validity of this method was examined both on the numerically synthesized data and the experimentally measured data for the phantom specimen and actual human subject. It was shown that an abdominal tomographic sound speed image could be successfully obtained by preparing only 32 transducer locations at the circumference around the abdominal surface and their combination of less than 100 number of observation path data as well as by avoiding the data intersecting the spinal cord. In addition, fat regions were extracted having a sound speed lower than the threshold value to demonstrate the possibility of this method for metabolic syndrome diagnosis.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.