Abstract

We used a microcomputer-controlled total ozone portable spectroradiometer instrument #21 (MTOPS21) to measure solar radiation at 298, 304 and 310 nm in Greenbelt, Md., during 1995. One day's ozone measurements from a Brewer instrument (B105) were used to calibrate the 304- and 310-nm channel ratios to a theoretical model. Total ozone estimates were then determined for the entire MTOPS21 data set. Differences between individual B105 and MTOPS21 ozone estimates show a 1% drop as solar zenith angles increase and depend on atmospheric attenuation and SO(2) variation at the ±2% level. Daily average values agree well (<0.5% average offset, 2% standard deviation).

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.