Abstract

To describe a method for assessing pressure variation vs. time (dp/dt) using blood flow acceleration measured by MRI, and to demonstrate its applicability in estimating left ventricular (LV) function. The method was tested in vitro using a pulsatile phantom, and a strong correlation was found between transducer and MRI determinations of dp/dt (r = 0.98). Selected aortic flow parameters were then measured in 10 patients and the results were compared with transducer measurements of the LV dp/dt. The correlation coefficients for the reference estimations of global myocardial function and MRI were 0.59 for aortic velocity, 0.74 for aortic acceleration, and 0.86 for aortic dp/dt. MR measurements of velocity and acceleration within the ascending aorta offer a noninvasive method for determining indices, such as the aortic dp/dt, that are closely correlated with the global myocardial contractility function.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.