Abstract

The leakage rate of air from a fume hood is an important factor for calculating the internal exposure at a workplace using the dose conversion factors recommended by the International Atomic Energy Agency. In this study, the leakage rate from a fume hood was estimated in a radioisotope laboratory using computational fluid dynamics simulations. For this purpose, a three-dimensional fume hood model was built using CAD software. In this model, the leakage rate was calculated to be 3%, with face velocities between 0.1 and 1.0 m/s at the sash opening of the fume hood. This rate rapidly decreased to zero as the face velocities decreased to less than 0.1 m/s.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.