Abstract
We evaluated the accuracy with which the Hologic QDR-4500A and QDR-2000 densitometers measure fat mass (FM) in 95 children. FM was derived from total body water measured by deuterium dilution (DD) in all children, by QDR-4500A in 50, and by QDR-2000 in 45 children. Compared with DD, both instruments underestimated FM (QDR-4500A: 3.35 +/- 2.5 kg, P < 0.0001; QDR-2000: 1.05 +/- 1.5 kg, P < 0.0001). Both absorptiometers showed magnitude biases relative to DD (QDR-4500: r = +0.70, P < 0.001; QDR-2000: r = -0.51, P < 0.001). We conclude that neither densitometer is equivalent to DD for estimation of children's FM. The QDR-4500A's current calibration seems to provide an even greater underestimate of FM than the QDR-2000.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Journal of Pediatric Gastroenterology and Nutrition
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.