A 6- to 7-MeV high-energy gamma-ray field, produced by the nuclear reaction of (19)F(p, αγ)(16)O, has been established at the Facility of Radiation Standards (FRS) in Japan Atomic Energy Agency for calibration purposes. Basic dosimetric quantities (i.e. averaged gamma-ray energy, air-kerma-to-dose equivalent conversion coefficients and air kerma rates at the point of test) have been precisely determined through a series of measurements using the NaI(Tl) spectrometer and an ionisation chamber coupled with an appropriate build-up material. The measurements obtained comply with values recommended by the International Organization for Standardization for an 'R-F field'. The neutron contamination component for the field has also been measured by means of a conventional neutron dose equivalent meter (the so-called neutron rem-counter) and determined to be ∼ 0.5 % of the total dose equivalent.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.