Abstract

The measurement of free intracellular magnesium (Mg2+) using the 31P chemical shifts of ATP requires the use of appropriate calibration solutions to determine the chemical-shift limits delta ATP alpha beta and delta MgATP alpha beta. Solutions containing excess Mg2+ contain significant amounts of Mg2ATP and yield positive errors in the value of delta MgATP alpha beta. For physiological applications this may overestimate free intracellular Mg2+ by as much as 300%. This error may be minimized if appropriate mole ratios of Mg2+/ATP are used to calibrate delta MgATP alpha beta.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.