Abstract

Accurate measurement of beam quality is very important in different areas of laser applications. Take the far-field Gaussian spot as an example, the calculation error formulas of peak Strehl ratio and beam quality factor β are derived in the present of various error sources, such as background photon noise, CCD pixel's non-ideal sampling, CCD dark voltage, read-out noise of CCD, signal photon noise and measurement window size. The calculation errors of laser beam quality changing with various error sources are presented in theory and the corresponding numerically simulated results are given. Theoretical and simulated results show that when the threshold is set at NB¯+σB, with NB¯ standing for average value of the error sources and σB standing for mean-square value of the fluctuation in error sources, the lowest calculating error of laser beam quality could be achieved.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.