Abstract

Tissue segmentation in CT images is a key parameter for accurate dose calculation in external beam radiation therapy. Tissue segmentation methods used in clinical practice cannot be applied to preclinical practice due to both CT energy (40kVp) and treatment beam energy (100- 300kVp). At MV energy range, tissue segmentation is based on the conversion of HU into densities (HU, ρ). At medium energy range, this conversion method does not allow to achieve a satisfactory accuracy. Taking into account tissue heterogeneities requires to define the tissue elemental composition (EC). We developed a new tissue segmentation method (HU, ρ, EC) allowing the automatic assignment of density and elemental composition from HU. The (HU, ρ, EC) method is dedicated to preclinical practice to reach accurate dose distributions in small animals.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.