Abstract

The application of inductively coupled plasma atomic emission spectrometry (ICP-AES) to the forensic sample was studied. On-line collection/concentration method of the sample with chelating resin column (TE-05) and air-sandwiched method to isolate the analyte zone at the highest concentration of an eluent was coupled with ICP-AES. The limits of detection (LODs) were much improved to 35 fold (Co): from Sc, 0.15 ng ml −1 to Ni, 1.99 ng ml −1, and the concentration efficiency was 7–14 times. This method was applied to the concrete with about 10 mg of samples. Major elements (Al, Fe, Mg) measured by conventional/ICP-AES and trace elements measured by this method, such as Cd, Co, Ni, and Pb were determined without matrices interference. Four concrete samples can be discriminated by comparing the content profiles of the trace elements and the major elements.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.