Abstract

Several factors have been shown to be associated with early development of radiolucent lines at the bone-cement interface in total knee arthroplasty (TKA). The posterior condylar surfaces, in particular, seem subject to poor cement technique, which could lead to early loosening. This study compares two cementation techniques in TKA, with respect to depth of cement penetration and radiolucency in the posterior condyles. All penetration depths were greater in group I (injected) versus group II (noninjected). Sixty-seven percent of group I showed penetration depths >1.5 mm compared with 23% of group II. No specimen in group I had gaps in the cement mantle on visual inspection or radiolucency on radiographic evaluation. No statistical differences, however, could be demonstrated between the two groups.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.