Abstract

In most cases, high energy solid state lasers such as Nd:YAG are used as source of illumination for Photoacoustic Tomography (PAT). However the bulkiness, high cost and low pulse repetition frequency (PRF) poses a challenge in translating this technology to an affordable clinical imaging option at bed-side. Pulsed Laser Diodes (PLD) on the other hand is portable, inexpensive and offers high PRF. However, the achievable depth of penetration using PLD is much lower than the solid state lasers. In this work, we demonstrate the feasibility of using sub-pitch translation approach on the receive-side ultrasound transducer to increase the depth sensitivity in PAT imaging system while using PLD as a source of illumination. The preliminary results obtained from experiments suggest that the higher density data obtained by augmenting raw RF lines from λ/2 positions of a linear array transducer provides better signal strength from deeper located targets and thereby increasing the depth of penetration by about 15% that reaches up to a depth of 14.3 mm.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.