Abstract

Food borne pathogens threaten food safety and affect human health. The lateral flow immunoassays (LFIAs) are widely concerned because of simplicity, low cost and user friendliness, and have broad application prospects in pathogen detection. However, the sensitivity of LFIAs is limited. Herein, multi-line LFIAs are introduced into pathogen detection for the first time. Compared with traditional single-line LFIAs, the overall signal strength of multi-line LFIAs has been significantly improved. It is particularly noteworthy that multi-line LFIAs detection accuracy of 103 CFU/mL pathogen has been improved by about 55%. The proposed multi-line LFIAs reduce the possibility of judging a positive result as a false negative result. The LFIAs strip was validated in real samples of milk and orange juice. This strategy has great potential for rapid detection of pathogens in real samples, and provides new insights for improving the accuracy and sensitivity of LFIAs strips.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.