Abstract

The effect of AC poling (ACP) on the piezoelectric and dielectric properties of diced sliver-mode vibrators (L13 × W0.15–0.30 × T0.51 mm3) was investigated for a convex abdomen ultrasound (US) probe that is made of a large Pb(In1/2Nb1/2)O3-Pb(Mg1/3Nb2/3)O3−0.3PbTiO3 (PIN-PMN-PT) single crystal (SC) plate (L13 × W60 × T0.51 mm3). The SC was manufactured by the continuous feeding Bridgman (CF-BM) method. The /ε 0 aging rate (AR) of both DCP and ACP large plate SCs were found to be within 1%/decade suggesting no obvious property degradation. After array dicing, a considerable decrease in electromechanical and dielectric properties was observed. After a post-dicing DCP, the diced DCP-DCP 0.15 mm wide SC slivers showed /ε 0 and electromechanical coupling factor (k′33) of 4300 and 89.1%, respectively. The diced ACP-DCP 0.15 mm wide SC slivers showed improved /ε 0 and k′33 of 5980 and 92.2%, respectively, which are considered useful in practical medical array transducer applications.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.