Abstract

When dealing with small amounts of 10Be, it is challenging to achieve the good performance in the ion source that is needed to enable a reliable AMS measurement. We have co-precipitated beryllium with dissolved NbCl 5, using a range of mixing ratios, to establish an optimal and practical procedure. 9Be 16O − beam currents for co-precipitated samples of 9Be–Nb are often higher, but show a slower rise than for samples either co-precipitated or mixed with silver. However, while the Ag sample currents soon deteriorate, Nb sample currents stay high for an extended time, leading to higher measuring performance and better statistics. Although co-precipitated Be–Nb mixtures do not reach such high beam currents as BeO mechanically mixed with Nb powder, we prefer co-precipitation as it provides enough material for handling and an additional mixing step is avoided. We recommend use of NbCl 5 as an alternative to AgNO 3 on account of the much lower boron association of niobium compared to silver. No critical limit in the matrix:Be mixing ratio was observed, but we note a steady decline of maximum currents reached with increasing mixing ratio.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.