Presents the results of independent energy determinations, one by ion-chamber measurements in a water phantom, one by film analysis in a lucite phantom, and one by the ion-chamber response ratio described by the authors. The energy differences are insignificant with regard to the corresponding mass stopping power ratios needed for calibration. The greatest variation exists for 6 MeV electrons, mainly a result of the steep slope on the descending portion of the ionisation-depth curve. Even in the worse-case situation, however, the differences in the stopping power values (L/ rho ) are less than 1%.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.