Abstract

AbstractThe purpose of the study was examination of energy dependence of Gafchromic EBT – a new radiochromic dosimetry film. The film response was measured irradiating the samples with doses from 0.1 Gy to 12 Gy using photon beams from the energy range 1.25 MV-25 MV. The samples were scanned and the film response was compared for different beam energies. The measurements were performed in order to evaluate potential use of radiochromic films in pretreatment dosimetric verifications of intensity modulated radiotherapy (IMRT) plans. A high uncertainty in readout of the film response was observed for films irradiated with doses lower than 1 Gy. The relative difference exceeds 20 % for doses below 1 Gy while for doses over 1 Gy the measured film response differs by less than 5 % for the whole examined energy range. No energy dependence of Gafchromic EBT film response was observed in the investigated energy range but the doses measured for pretreatment plan verifications should exceed 1 Gy.Keywordsfilm dosimetryradiochromic films

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.