We report a pyrometer that operates at a mid-infrared wavelength range (7–8 μm), where sapphire substrates are opaque and the reactant gases are transparent. The pyrometer also employs a novel “self-reflectance” method of emissivity correction, whereby thermal emission from the sample serves as a radiation source to measure its emissivity. The instrument was installed on a multiwafer Veeco D-125 MOCVD system and used to measure the emissivity-corrected temperature at a variety of GaN and InGaN deposition conditions. For a series of InGaN multiquantum well growth, the pyrometer was used to control the surface temperature at the critical InGaN deposition step, resulting in improved control of the photoluminescence wavelength.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.