Abstract

Abstract Previous studies have reported a significant residual signal in LiF:Cu,Mg,P (GR-200) which arises from a peak at 270 oC with an intensity of 7-50% relative to the main dosimetry peak. A method has been found to prepare LiF:Cu,Mg,P in two forms; one with a TL sensitivity 30-40 times greater than TLD-100 but with a residual signal of only 2-3% following 240 oC for 10 s readout, the other, even more optimised material (SSD-NR), with a sensitivity 15-20 times that of TLD-100 but with the 270 oC peak almost totally eliminated, leading to a residual signal of 0.25-0.5%.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.