Abstract

An accurate method for determining elemental analysis using gamma–gamma coincidence counting is presented. To demonstrate the feasibility of this method for PGNAA, a system of three radioisotopes (Na-24, Co-60 and Cs-134) that emit coincident gamma rays was used. Two HPGe detectors were connected to a system that allowed both singles and coincidences to be collected simultaneously. A known mixture of the three radioisotopes was used and data was deliberately collected at relatively high counting rates to determine the effect of pulse pile-up distortion. The results obtained, with the library least-squares analysis, of both the normal and coincidence counting are presented and compared to the known amounts. The coincidence results are shown to give much better accuracy. It appears that in addition to the expected advantage of reduced background, the coincidence approach is considerably more resistant to pulse pile-up distortion.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.