Abstract

An electrothermal vaporization device has been designed for use in sample introduction and this has been coupled with microwave-induced plasma atomic absorption spectrometry (ET-MIP-AAS). Various vaporization chambers were tested to minimize the turbulence and the dead volume. A series of experimental parameters were examined and optimized. Some influences, such as concomitant element and easily ionized element (EIE) effects, which are commonly encountered in microwave-induced plasma atomic emission spectrometry were investigated. The analytical figures of merit for ET-MIP-AAS are discussed. The characteristic concentrations obtained in this work were found to be at the ng ml–1 or higher level. The precision for the determination of Ca, Ag and Cu was shown to be in the range 5.6–7.8%. The recommended method was also used to analyse real samples, and satisfactory results were obtained.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.