Abstract

Ultrasonic phased or sector array transducers were built using PMN-PT based electrostrictive materials. Large field-induced piezoelectric coefficients were obtained in these relaxor-type materials at relatively low DC bias (K/sub t/=0.43-0.50 at 5 KV/cm). The electrostrictive transducers have similar sensitivity and bandwidth compared to conventional PZT transducers. By choosing the appropriate composition and process, the sensitivity variation of electrostrictive transducers can be minimized to within 3 dB over the operating temperature range. The clinical results demonstrate the benefit of using electrostrictive materials to obtain elevation aperture control.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.