Abstract

Three electronic personal dosimeters (EPD-N) manufactured by Siemens, serial numbers 0635, 0658, and 0683, were tested at the Radiation Calibration Laboratory for an evaluation of their response to neutron, gamma and x-ray radiation. Designed to provide real-time neutron and photon dosimetry, the EPD-N is capable of estimating and displaying neutron and gamma dose components separately for a range of energies from 50 keV to 7 MeV for photon and 0.025 eV to > 10 MeV for neutron. All tests were conducted using the factory calibrations. A technical representative of the manufacturer indicated that site-specific calibrations are required as factory settings are calibrated for the lowest neutron energy limit of 0.025 eV. This raises concerns about the reliability of these devices in measuring neutrons when calibrations are made for a specific site radiological characterization then used at another site.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.