Abstract
We have used household consumables to facilitate electrochemical etching of stainless-steel hypodermic tubing to produce tapered-tip emitters suitable for electrospray ionization for use in mass spectrometry. The process involves the use of 1% oxalic acid and a 5 W USB power adapter, commonly known as a phone charger. Further, our method avoids the otherwise commonly used strong acids that entail chemical hazards: concentrated HNO3 for etching stainless steel, or concentrated HF for etching fused silica. Hence, we here provide a convenient and self-inhibiting procedure with minimal chemical hazards to manufacture tapered-tip stainless-steel emitters. We show its performance in metabolomic analysis with CE-MS of a tissue homogenate where the metabolites acetylcarnitine, arginine, carnitine, creatine, homocarnosine, and valerylcarnitine were identified, all with basepeak separated electropherograms, within <6 min of separation. The mass spectrometry data are freely available through the MetaboLight public data repository via access number MTBLS7230.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.